FAQ

Click plus (+) button to see answer

How do PurGrip hex slot implant drivers work?

The working end of the PurGrip hex slot implant driver tool incorporates a unique “spring clip” design feature  with a solid hex core at the very tip of the driver shaft. This unique patented design allows for the “decoupling” of the gripping action from the torqueing action of the PurGrip hex slot implant driver tool and prevents the spring clip feature from losing its “spring” grip each time you use it.

Does the length or diameter of the screw you want to place or remove matter?

No. The only dimension that matters is the size of the diameter hex slot of the head of the screw. The current PurGrip hex slot implant driver tools are made to grip screw head hex slots of  1.27mmØ.

What is the maximum torque and speed (RPM) for PurGrip hex slot implant driver use?

Do not exceed 25 Ncm torque (for both hand and rotary PurGrip tools) and 80 RPM (for rotary PurGrip tool only). If you exceed the recommended torque force you can permanently damage the PurGrip hex slot implant driver tool. Discard PurGrip hex slot implant driver tool if damaged.

How do I know if the PurGrip hex slot implant driver tool is damaged?

If the working end of the PurGrip hex slot implant driver tool when fully engaged in the hex slot of the screw head no longer grips the screw securely, the PurGrip hex slot implant driver tool is damaged and should be discarded immediately.

If I want to tighten screw to more than 25Ncm torque what should I do?

Insert and thread screw in with PurGrip hex slot implant driver tool and then finish tightening to desired torque using standard latch hex driver tool.

Can I pick up a screw directly from an implant package without handling the screw?

Yes. Make sure to fully engage the working end of the PurGrip hex slot implant driver tool in the hex slot of the screw before removing the hex slot screw from packaging.

How do I know when the PurGrip hex slot implant driver tool is fully engaged?

Press the working end of the PurGrip hex slot implant driver tool down as far as you can into the hex slot of the screw. When it is fully engaged in the hex slot, you will hear a soft “click” sound that indicates the “spring clip” feature of the PurGrip hex slot implant driver tool is fully engaged and gripping the internal walls of the screw head’s hex slot.

How to I reduce the risk of the PurGrip Hand hex slot implant Driver tool itself from being aspirated?

Tie a sufficient length of floss around the indented circumferential slot on the distal (non-working end) of the knurled tool handle before use.

How do I clean & sterilize the PurGrip hex slot implant drivers between patients?

The IFU contains directions on cleaning and sterilization of your PurGrip hex slot implant driver tools which are standard for all dental surgical tools.

How often do I need to clean and sterilize PurGrip hex slot implant driver tools?

Clean & sterilize after each use.

How do I disengage the PurGrip hex slot implant driver tool from the screw?

Once you have fully inserted the screw by threading it in with the PurGrip hex slot implant driver tool, simply lift the working hex end of the PurGrip hex slot implant driver tool from the hex slot of the screw to disengage the PurGrip hex slot implant driver tool from the screw.

How do I engage the PurGrip Rotary hex slot implant driver tool into the handpiece of a Physio unit?

Insert the slotted latch end of the PurGrip hex slot implant driver tool into the head of the handpiece and engage the internal or external latch mechanism to secure the driver shaft in the handpiece. Make sure to verify that the slotted latch end is secure by tugging on the working end of the PurGrip hex slot implant driver tool.